Integrated input |
Input |
2CH (no isolation btw. CH)
Dry contact or Open collector (NPN)
Contact rate: 5 V DC, 10 mA |
Count range |
0 to 99999999 (Start from 0 when an integration value exceed 99999999) |
Input frequency |
HF: 1250 Hz Max. Pulse width more than 400 µs
LF: 100 Hz Max. Pulse width more than 5 ms |
Pulse parameter |
×0.001, ×0.005, ×0.01, ×0.05, ×0.1, ×0.5, ×1, ×5, ×10, ×50, ×100 |
Computation |
A+B, A-B for Integrated input A, B |
Reset |
RESET or Start printing (Hourly report, Daily report, Monthly report, and Monthly report memory). |
Work time |
Input |
2 points.
Dry contact or Open collector (NPN)
Contact rate: 5 V DC. 10 mA |
Count range |
0 to 536870911s (approx. 17 years) |
Reset |
RESET or Start printing (Hourly report, Daily report, Monthly report, and Monthly report memory). |
Indication |
Integrated |
0 to (99)999999 Last 6 digits, with Zero suppress. When the lower 6 digits exceed 999999, the decimal point at 106 is flashing. Cycle: 100ms
Integration A, Integration B, Integration computation (A+B, A-B), Switching indication (Date - Integration). |
Clock |
Year/Month/Date or Hour/Minute/Second |
Control Signal |
Input |
-
MANUAL, ADJ, RESET
-
Dry contact input or Open collector (NPN).
Contact rating 5 V DC. 10 mA.
MANUAL, RESET: Pulse width more than 10ms.
ADJ: Pulse width more than 1s.
|
Output |
-
PE Transistor ON at paper end.
-
Open collector (NPN) 30 V DC. / 30 mA, Max. Voltage less than 1.6 V
|
Print Style |
Thermal line dot. |
Character |
Alphabet, Numbers, Katakana, Symbols, etc. |
Dot |
16 × 16(2 mm × 2 mm) |
Digit |
24 digits, Max. |
Printing speed |
Approx. 22.5 mm/s, 6 lines/s, Max. Note) Printing rate: Less than 16%. |
Printing width |
46 mm |
Life time |
At 25 °C
-
Head:
-
More than 109 pulse (pulse resistance)
50 km or longer except damage by foreign particle, alien substance.
(abrasion resistance)
|
Paper |
-
58 mm width x 48φ (inside diameter 12φ)
-
Length 25 m (approx. 6500 lines printable)
Use specified chart paper, otherwise printing quality and products lifetime will be out of warrantee.
|
Calendar Clock |
Accuracy: ±3s, per day. (at 25°C)
A leap year adjustment: Automatic adjustment till 2099. |
Power failure measure |
Set value, Integration value, Work time, Monthly reports are memorized by Non-volatilization memory. |
Power Supply |
AC 100 to 240 V 50/60 Hz DC 24 V ± 10% |
Power Range |
AC 90 to 250 V DC 21.6 to 26.4 V |
Power Consumption |
AC 100 V:16 VA, AC 200 V: 20 VA
DC 24 V: 500 mA |
Operating temperature & humidity |
0 to 50°C, less than 85% RH (no condensation) |
Storage Temperature |
-20 to 60°C |
Weight |
Approx. 700 g |
Installation |
Panel mounting |
Withstand Voltage |
-
Input / Output - Power:
-
1500 V AC. For 1min. (AC powered)
500 V AC. For 1min. (DC powered)
|
Insulation Resistance |
-
Input / Output - Power:
-
500 V DC. More than 50 MΩ (AC powered)
500 V DC. More than 50 MΩ (DC powered)
|
Option |
Analog output DC 1 to 5 V or DC 4 to 20 mA |