Item Code | 212908 |
---|---|
Model | DS611 |
Circulation Type | Natural Convection |
Operating Temp. | Room Temp. +5 to 260deg.C |
Temp. Adjustment Accuracy | ±1deg.C (at 260deg.C) |
Temp. fluctuation | ±1.5deg.C (at 260deg.C) |
Temp. Distribution Accuracy | ±5deg.C (at 260deg.C) |
Temp. gradient | 20deg.C (at 260deg.C) |
Max. Temp. Reaching Time | Approx. 90 min |
Interior Material | Stainless Steel |
Exterior Material | Cold Rolled Steel Plate with Melamine Resin Baked Finish |
Heat Insulating Material | Glass Wool |
Heater | SUS Pipe heater 1.36kW |
Sensor | K-Thermocouple (for temp. adjustment and overheat prevention) |
Cable Hole | I.D. 33mm (Right Side) |
Exhaust Port | I.D. 33mm x 2 (Top Side) |
Observation Window | 250 x 280mm Tempered Glass x 2 pcs |
Temp. Control System | PID Control by Micro Computer |
Temp. Setting | Digital Setting by UP/DOWN Key |
Temp. Display | Digital Display by Green LED |
Other Display | Temp. Pattern Display for Operating Display by LED |
Timer | 1min. to 9hrs.59min. or 100hrs. to 999hrs. |
Timer Resolution | 1min. or 1hrs. |
Operating Function | Fixed Temp. Operation, Automatic Start, Automatic Stop, Program Operation , Quick automatic stop |
Additional Function | Timer Function (Actual Timer within 24hrs.), Total Operation Hours Timer (to 65535hrs.), calendar time ( 24 hours), calibration offset, Power discharge, CO2 discharge, power failure return mode, user configuration information |
Heater Circuit Control | Triac Zero-cross Type |
Safety Countermeasure | Self Diagnosis Function (Abnormal Temp. Sensor, Heater Wire Broken, Triac Short Circuit, Automatic Overheat Prevention), Independent Overheating Prevention, Key Lock, Electric Leakage Breaker |
Internal dimensions(W×D×H mm) | 600×540×500mm |
External dimensions(W×D×H mm) | 710×650×870mm |
Power Source |
AC220V 6.8A Single phase |
Weight | 64kg |