Specification
Measurement principle
|
Luminescence
|
|
|
||
Parameter
|
Dissolved Oxygen
|
|
Measurement range
|
0...20 mg/L
0...20 ppm
0...200 %
|
|
Measurement accuracy
|
± 0.1 mg/L
± 0.1 ppm
± 1 %
|
|
Resolution
|
0.01
|
|
Measurement interval
|
> 5 s
|
|
Resolution (temperature)
|
0.01 °C
|
|
Accuracy (temperature)
|
0.5 °C
|
|
Membrane Cap
|
No cross-sensitivity with : pH 1 – 14 ; CO2, H2S, SO2
Cross-sensitivity to Organic solvents, such as acetone, toluene, chloroform or methylene
chloride Chlorine gas
|
|
|
||
Material
|
Standard Version in passivated Stainless steel 316L body, strainer and screw,
For Seawater application Version in Titanium body, strainer and screw,
Cable : polyurethane jacket
Steam gland : Polyamide
Patch with active material (black) – Membrane : Optical isolation silicon
|
|
Dimensions (L x Ø)
|
146 mm x 25 mm
|
~ 5.7Ý x 1Ý
|
Weight stainless steel
|
~ 450 g
|
~ 1 lbs
|
Weight titanium
|
~ 300 g
|
~ 0.7 lbs
|
|
||
Interface
|
RS-485 (Modbus RTU)
|
|
Power consumption
|
1 W
|
|
Power supply
|
12 V (± 10 %)
|
|
Sensor cable
|
2 m and 10 m
|
|
|
||
Calibration/maintenance interval
|
2 years
|
|
Warranty
|
1 Year (EU & US: 2 years) on electronics; All wearing parts are not included in the warranty
|
|
|
||
Max. pressure
|
5 bar
|
~ 72.5 psig
|
Protection type
|
IP68
|
NEMA 6P
|
|
||
Sample temperature
|
0...+50 °C
|
~ +32 °F... +122 °F
|
Ambient temperature
|
0...+50 °C
|
~ +32 °F... +122 °F
|
Storage temperature
|
-10...+60 °C
|
~ +14 °F... +140 °F
|
Versions