Code | Ideal For | Approx. Thickness - Block 1 | Approx. Thickness - Block 2 | Approx. Thickness - Block 3 | Approx. Thickness - Block 4 | Accuracy |
P1 |
PosiTector 6000 FT, FTS, NTS, FNTS, FTRS PosiTector 200 D |
375 μm 15 mils |
2 mm 80 mils |
4.5 mm 185 mils |
6.5 mm 250 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P2 | PosiTector 6000 FHS, FJS, NHS, EOC | 2.5 mm 100 mils |
6.5 mm 250 mils |
13 mm 500 mils |
19 mm 750 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P3 | PosiTectAor 100 C | 375 μm 15 mils |
1.5 mm 60 mils |
2.5 mm 100 mils |
4.5 mm 185 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P4 | PosiTector 100 D | 1.5 mm 60 mils |
2.5 mm 100 mils |
4.5 mm 185 mils |
6.5 mm 250 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P5 | PosiTector 6000 FKS, NKS | 1.5 mm 60 mils |
2.5 mm 100 mils |
6.5 mm 250 mils |
12 mm 480 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P6 | PosiTector 200 C | 375 μm 15 mils |
1.5 mm 60 mils |
2.5 mm 100 mils |
3 mm 125 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P7 | PosiTector 6000 FHXS | 1.5 mm 60 mils |
4.5 mm 185 mils |
6.5 mm 250 mils |
9.5 mm 375 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P8 | PosiTector 6000 FNGS, FLS | 13 mm 500 mils |
13 mm 500 mils |
13 mm 500 mils |
19 mm 750 mils |
± (2.5 μm + 0.05% of thickness) ± (0.1 mil + 0.05% of thickness) |
P1-P7 Blocks: 38 x 70 mm (1.5'' x 2.75'') P8 Blocks: 76 x 76 mm (3.0'' x 3.0'')
Certified coated metal plates and polystyrene blocks are ideal for verifying the accuracy and operation of coating thickness gages and are an important component in fulfilling both ISO and in-house quality control requirements.
Many organizations require verification of gage accuracy at the test site each time a coating thickness gage is put into service and at frequent intervals during use. Ideal for this purpose, DeFelsko certified coating thickness standards have measured values traceable to a National Metrology Institution.
- Used to verify the accuracy and operation of any Type 1 (mechanical) and Type 2 (electronic) magnetic, eddy-current or ultrasonic coating thickness gage
- Ideal for use in the calibration lab, in the field or on the factory floor
- Standards with steel or aluminum substrates consist of 4 plates mounted in a protective binder
- Polystyrene thickness standards consist of 4 blocks supplied in a rugged acrylic storage box
- Individually serialized for traceability to NIST or PTB - includes a Certificate of Calibration
- Certified and labeled in both Metric and Imperial units