- Five optional sensors for measuring special shaped work-pieces.
- DSP chip controlled achieves high accuracy and fast testing speed.
- Matrix LCD screen displays all parameters and data.
- 13 parameters and 4 filtering methods: RC, PC-RC, GAUSS and D-P.
- High quality with ISO/ DIN/ CE/ JIS/ FCC standards.
- Memory: 100 groups of measuring data.
- Power supply: 3.7V rechargeable Li-ion battery.
- Standard Sensor: diamond stylus with radius 5μm, force 4mN(0.4gf), and angle 90°
- Available PC software for data management and Excel reports.
- Optional printer connecting by USB.
Standard Sensor
Detection Principle | Material of Stylus | Radius of Stylus | Stylus Force | Contact Angle |
Inductive | Diamond | 5μm | 4mN(0.4gf) | 90° |
Technical Specification
Model No. | Leeb432 |
Roughness parameters | Ra、Ry、Rq、Rz、Rt、RSm、RS、Rp、Rv、R3z、RSk、Rmax、Rmr |
Measuring range | Ra:0.005-16μm Rz:0.02-160μm |
Resolution | 0.001μm |
Maximum driving trip | 17.5mm/0.7inch |
Sample length/Range (L) | 0.25mm, 0.8mm, 2.5mm/±20μm,±40μm,±80μm |
Evaluation length | 1.25mm, 4mm,5mm/Option from(1-5)L |
Filtering methods | RC, PC-RC, GAUSS, D-P, |
Accuracy | ≤±8% |
Repeat ability | <6% |
Memory | 100 Groups |
Operating temperature | 0 ºC -40 ºC |
Weight | 440g |
Dimensions | 119×47×65mm |
Standard configuration | Main Unit,Standard sensor,Calibration block, Supporting platform, Power charger, Manual,Certificate, Warranty card, PC analysis software |
Optional Accessories | Measurement Platforms , optional sensors, printer |